N4 removes the slow intensity variation that coil sensitivity puts on MRI, in this browser, with ANTs N4BiasFieldCorrection compiled to WebAssembly (the same ANTs commit as SYNcro; its bias field equals native ANTs 2.6.5). For a DWI with its bval and bvec it follows MRtrix dwibiascorrect ants: N4 on the mean b0 weighted by the brain mask, the field rescaled to keep the brain's total signal, and every volume divided by it. A 3D image is corrected directly. The settings are dwibiascorrect's: shrink factor 4, B-spline mesh 100 mm of order 3, up to 1000 iterations. In a diffusion pipeline, correct the bias field after motion and distortion correction. A dropped BIDS folder gives each DWI run with its bval, bvec and inherited JSON sidecar; choose the run when there is more than one.